User Manual MAS-100 Regulus en.docx, Version: 4.0
13) Select the perforated lid of the sampling head:
300×0.6 or 400×0.7
14) Choose the calibration validity:
Values from 0 to 12months may be entered.
A value of 0 disables the calibration validity
check upon starting up the air sampler.
Press Set calibration validity to set the value.
15) Set the optional calibration process parameters:
Measurement blocking upon expired calibration.
Acoustic feedback for the Run time check.
16) Set the value for the Maximum allowed deviation.
Note
The standard factory setting for the Calibration validity is 12months.
Note
The standard factory setting for the Maximum allowed deviation is ±2.5%.
17) Later in the process, the PC Software creates auto-
matically an Adjustment and calibration certificate
and saves it to a predetermined destination.
To customize the file destination, click the but-
ton and define the file path.
18) In addition, the Adjustment and calibration certifi-
cate will be printed automatically if the Print certifi-
cate box is ticked.
19) Select Automatic adj. + calib with certificate.
20) Set the adjustment parameters:
Adjustment Flow
Adjustment Span
Start the process with the button.
Pressing afterwards aborts the process.
Note
The standard factory setting for the adjustment Flow is 100SLPM.
Note
The standard factory setting for the adjustment Span is 10SLPM.