User Manual MAS-100 Regulus en.docx, Version: 4.0
21) Press Yes to proceed.
Do not add or remove instruments.
Pressing No aborts the process.
22) The user will be asked whether to activate the cali-
bration validity immediately.
Press Yes if the calibration validity duration shall
be set starting from now.
Press No if the calibration validity duration shall
be set later, i.e. when the air sampler is turned
on the next time.
23) The air sampler will now adjust through all three adjustment points. No action from
the user is required, the process will continue automatically..
24) Once the process is done, the final calibration re-
sult will be displayed by the value Deviation from
the Target Air Flow.
A calibration was successful if:
Deviation from the Target Air Flow is within ± the
Maximum allowed deviation.
Successful calibration, value
marked green:
Failed calibration, value marked
red:
Note
The calibration also fails if at least one of the values in the row of Devia-
tion from target air flow is out of the allowable range.
25) Upon a successful adjustment:
- The adjustment values are stored in the air sampler
- The Adj. and calibration certificate is exported automatically to a preset file desti-
nation
- The Adj. and calibration certificate is printed (if function is activated)
26) If desired, the user may create an own certificate:
Press Adjustment and calibration certificate.
Enter calibration certificate number.
Press OK.