Page 123 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
15) Iteration:
- Modify Blower PWM
- Take anemometer reading
- Is the F
m
value firmly inside the range of Adjustment flow ±0.5 SLPM?
o If No: Redo iteration.
If Yes: Press the Next button and continue.
Example
Adj. Point
❶ ❷ ❸
Adj. flow 95.0 SLPM 105.0 SLPM 100.0 SLPM
Tolerance ±0.5 SLPM
Adj. flow range 94.5 - 95.5 SLPM 104.5 - 105.5 SLPM 99.5 - 100.5 SLPM
16) The message Waiting will briefly appear.
No action from the user is required, the process will
continue automatically.
Pressing Stop interrupts the process and reverts to
the Adj./Cal. Mode menu.
17) The message Waiting will briefly appear.
No action from the user is required, the process will
continue automatically.
Pressing Abort interrupts the process and reverts
to the Adj./Cal. Mode menu.
18) Enter the last calculated mean flow value F
m
by
modifying the displayed value with the / but-
tons.
Press Next to save the value and to proceed.
Pressing Abort interrupts the process and reverts
to the Adj./Cal. Mode menu.
Adjustment point ❷
19) Proceed as with steps 13-18 with the adjustment flow rate corresponding to Adjust-
ment Point ❷.
Adjustment point ❸
20) Proceed as with steps 13-18 with the adjustment flow rate corresponding to Adjust-
ment Point ❸.
21) The message Waiting will briefly appear.
No action from the user is required, the process will
continue automatically.
Pressing Abort interrupts the process and reverts
to the Adj./Cal. Mode menu.