Page 136 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
Note
Upon a failed calibration, only the option Create test protocol is available.
18) Press Create test protocol.
19) Enter the the Test protocol Nr.,
then Press OK.
20) A preview of the certificate is displayed in a sepa-
rate window. Users may now modify the layout to
their needs.
Press the Print button to print a paper certificate,
and/or
Press the Save button to store the certificate as a
PDF file.
21) If used, remove the MAS-100 Family Calibration Disk or the Petri dish.
22) This procedure may be repeated for all other sampling heads.
23) End of process.