12) The instrument will proceed to the K
5
menu.
This parameter must be set to 0.
13) Modifying the K
5
value:
– Values 0 - 30 are possible
– After reaching 30, the counter restarts at 0.
Press the YES button to increase the value until the desired value is reached.
or
Press the NO button to save the current value and continue.
14) Press the YES button until the K
5
value reaches 0.
15) Press NO to continue.
16) The instrument is now in the calibration reminder
menu.
17) The user may enter now the desired number of
months for the calibration reminder.
Expired calibration warning
18) Enter the desired calibration validity duration by toggling with the YES button.
– Values from 0 to 12 months may be entered.
– After reaching the value of 12 the counter will restart with 0.
– A value of 0 disables the calibration validity check upon starting up the air sam-
pler.
– Once the desired value is entered, press NO to save the value and continue.
19) Press YES to finish the adjustment process and to
return to the main menu.
20) End of process.