User Manual MAS-100 Regulus en.docx, Version: 4.0
13) Press YES to proceed to the calibration reminder
menu.
or
Press NO to return to the K
3
menu.
14) After pressing YES the air sampler will continue
with the calibration reminder menu.
15) The user may enter now the desired number of
months for the calibration reminder.
Expired calibration warning
16) Enter the desired calibration validity duration by toggling with the NO button.
– Values from 0 to 12 months may be entered.
– After reaching the value of 12 the counter will restart with 0.
– A value of 0 disables the calibration validity check upon starting up the air sam-
pler.
– Once the desired value is entered, press YES to save the value and continue.
17) The adjustment process is now finished and the in-
strument will return to the main menu.
18) Remove the MAS-100 Family Calibration Disk or the Petri dish.
19) End of process.