101.0SLPM on the anemometer. The ambient temperature is 22°C and the ambient pressure
980mbar. Calculate the equivalent volume flow.
V
󰇗
=
V
󰇗
p
N
T
p T
N
=
101.0 SLPM 1013.25 mbar 295.15 K
980mbar 293.15K
= 105.1 L/min