Example
Adj. Point
❶
Adj. flow 100.0 SLPM
Tolerance ±0.5 SLPM
Adj. flow range 99.5 - 100.5 SLPM
19) Enter the calculated Mean flow value.
Finalize pressing the button.
20) Press the Yes button to accept the adjustment and
store the values in the air sampler.
Pressing No aborts the process.
21) If this process must be repeated for different perforated lids, start over on step 12 and
use the other perforated lid.
22) End of process.