Page 60 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
13) The process will now subsequently perform flow
measurements and display the result.
Pressing Stop interrupts the process and reverts to
the Method menu.
14)
A time check will be done, once all flow measure-
ments are done.
Pressing Stop interrupts the process and reverts to
the Method menu.
15) The calibration validity can now be activated.
Press Yes to set the calibration validity, or
Press No to interrupt the process and return to the
Method menu.
The air sampler must be re-adjusted according to
chapter 6.3, chapter 6.4 or chapter 6.6 if the cali-
bration result is outside the specified range.
16) Enter the desired calibration validity duration with
the / buttons:
– Values from 0 to 12 months may be entered.
– After reaching the value of 12, the counter will
restart with 0.
– A value of 0 disables the calibration validity
check upon starting up the air sampler.
Once the desired value is entered, press OK to save
the value and continue.
Note
The standard factory setting for the Calibration validity is 12 months.
17) The air sampler will return to the Method menu.
18) If this process must be repeated for different perforated lids, start over on step 10 and
use the other perforated lid.
19) End of process.