Page 85 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
11) Choose with the buttons Man adj.
Press the ► button to continue.
12) The blower starts up and the current rotation
speed is displayed on the air sampler.
13) Adjust the rotation speed with the buttons until the flow is stable and in the de-
sired range.
14) Anemometer reading:
– Allow the anemometer 40 seconds to settle prior taking readings.
– Take 3 consecutive readings from the anemometer.
– Note these 3 flow values: F
1
, F
2
, F
3
– Calculate the mean flow value: F
m
= (F
1
+ F
2
+ F
3
)/3
15) Iteration:
- Modify rotation speed
- Take anemometer reading
- Is the F
m
value firmly inside the range of 100 L/min ±0.5 L/min (99.5 - 100.5)?
o If No: Redo iteration.
If Yes: Press the
button and continue.
16) The air sampler stores the adjustment values.
Press the ► button to continue.
17) The air sampler will display the calibration validity.
Press the ► button to continue.
18) Press the button to get to the Exit ? option.