11) After successful login, the PC Software starts in the Adjustment/Calibration menu.
12) Select the perforated lid of the sampling head:
300 × 0.6 or 400 × 0.7
13) Select Manual 1-point.
14) Set the desired Adjustment flow.
Start the process with the button.
Pressing afterwards aborts the process.
Note
The standard factory setting for the Adjustment flow is 100 SLPM.
Note
The Adjustment span value has no influence in a Manual 1-point adjustment
15) Press Yes to proceed.
Do not add or remove instruments.
Pressing No aborts the process.