63Service manual for medifa 6000 mobile operating tables, version 2.0
Maintenance plan
11. Maintenance plan
Manual switch (cable/IR)
No. Feature
Test
equipment
Set value/Display Actual value/test result
1 Check housing Visual No cracks or other damage
2 Check cables Visual No crushing points or other damage
3 Check connection plug
and connection socket
Visual Test Probes OK, no contamination,
plug firmly seated in socket
4 Check membrane
keyboard
Visual,
Function
No damage, pictograms visible, all
functions can be carried out
5 Check display Visual,
Function
Display shows the correct execution
of the functions
Column
No. Feature
Test
equipment
Set value/Display Actual value/test result
1 Check membrane
keyboard
Visual,
Function
No damage, pictograms visible, all
functions can be carried out, Diodes?
2 Check cladding Visual No damage
3 Check bellows Visual No damage
4 Check IR receiver Visual No damage or soiling
5 Check lifting column Allen key Firm fit, backlash-free telescopic
segments
6 Check Trendelenburg
drive
Visual Fixed mounting, safe function,
smooth running
7 Check lateral drive Visual Fixed mounting, safe function,
smooth running
8 Check hydraulic system
for leaks
Visual No visible oil drops or leaks
9 Check cabling Visual No sharp bends or crushing points
10 Check type plate Visual No damage, good perceptibility