64 Service manual for medifa 6000 mobile operating tables, version 2.0
Chassis and base
No. Feature
Test
equipment
Set value/Display Actual value/test result
1 Check pictograms Visual No damage, good perceptibility
2 Check brakes Function The operating table is stable and
firm.
3 Check rollers Visual No damage
4 Check hydraulic system
for leaks
Visual No visible oil drops or leaks
5 Check free running of the
rollers
Function Easy to move
6 Check built-in plug of
mains connection
Visual No damage
7 Check mains cable Visual No sharp bends or crushing points
8 Check potential
equalisation flush plug
Visual No damage
9 Check potential
equalisation cable
Visual No sharp bends or crushing points
10 Check cabling Visual No sharp bends or crushing points
Top frame
No. Feature
Test
equipment
Set value/Display Actual value/test result
1 Check padding Visual No damage, no soiling
2 Check padding plates Visual No damage, firm fit
3 Check standard rails Allen key Firm fit
4 Check toothed belt Visual No damage
5 Sliding drive Visual Quiet operating
6 Grease spindle for
longitudinal adjustment
Bold,
Visual
Sufficient grease on the spindle
7 Check upper back drive Visual Fixed mounting, safe function,
smooth running
8 Check lower back drive Visual Fixed mounting, safe function,
smooth running
9 Check leg plate drives Visual Fixed mounting, safe function,
smooth running
10 Check hydraulic system
for leaks
Visual No visible oil drops or leaks
11 Check cabling Visual No sharp bends or crushing points
12 Check pictograms Visual No damage, good perceptibility