65Service manual for medifa 6000 mobile operating tables, version 2.0
Maintenance plan
Safety test according to VDE 0751
No. Feature
Test
equipment
Set value/Display Actual value/test result
1 Check connecting cable,
protective conductor,
labels and fuses
Visual No damage, no soiling
2 Check protective
conductor resistance
Test device < 0.3 Ω
3 Primary check of
insulation resistance
Test device > 2 MΩ
4 Check device leakage
current
Test device < 0.5 mA
5 Check leakage current for
patient
Test device < 5 mA