Delivery, Transport and Setting Up
19 D53182 02/2024
■ If the appliance has its own castors, also release the locking device on the (front) castors
4.4 Unpacking
■ Do not unpack the appliance until you reach the installation site.
■ Remove the cardboard packaging by pulling it upwards or carefully cutting along an edge
or unscrew and remove wooden crate.
Checking for completeness and transport damage
■ Check the delivery note to ensure the delivery is complete.
■ Check the unit for damage.
If you notice deviations from the scope of delivery, damage or anything unusual, do not put
the unit into operation and inform the haulage company and the manufacturer.
Removing the transportation lock
■ Remove the transportation lock. It is located between the door hinge, door and frame and
has to be removed after opening the door.
Disposing of packaging material
■ Dispose of the packaging material (cardboard, wood, foil) in accordance with the
applicable disposal regulations for the respective material in your country.
4.5 Storage after Delivery
If the unit is initially to be stored after delivery:
■ Observe storage conditions (see }10.1Storage and Transport)
See also
2 Storage and Transport [}56]
4.6 Setting Up
WARNING
Danger of tipping due to the appliance’s centre of gravity
Due to its centre of gravity, the appliance could tip forwards and injure you or someone
else.
– Always attach the appliance to a wall with the anti-tilt bracket.
– In case there is not enough space to fasten the appliance to a wall, do not put the
appliance into operation and do not open the door.
– Contact Memmert service.
4.6.1 Preconditions
ü The installation site must be flat and horizontal and must be able to reliably bear the
weight of the unit (see
}3.7Technical Data). Place the unit on a heat-resistant, fireproof and non-flammable
surface.
ü Depending on the model (see }3.6 Nameplate), a 230 V, 115 V or 400 V power
connection must be available at the installation site.
ü The distance between the wall and the rear panel of the appliance must be at least
15cm.