Menu Mode
50 D53182 02/2024
10. Wait until the appliance reaches and displays the setpoint temperature.
■ The reference instrument will display the corresponding deviation.
11. In the SETUP, adjust the calibration correction value Cal2 to the deviation temperature
(actual value measured minus setpoint value).
12. Save the setting by pressing the confirmation key.
13. Compare the temperature measured by the reference measurement instrument with
the temperature displayed on the appliance.
ð After the calibration procedure, the temperature measured by the reference instrument
should now also be the setpoint temperature.
With Cal1, a further calibration temperature below Cal2, and with Cal3 a temperature
above, can be programmed in the same manner. The minimum difference between the Cal
values is 20K.
i
If all calibration correction values are set to 0.0 K, the factory calibration settings are
restored.
8.6 Programme
In the Programme display, programmes created using the AtmoCONTROL software can be
transferred to the appliance and saved on a USB storage medium. Here, you can also
select the programme you wish to run (see }6.5.3 Programme Mode) and delete
programmes.
1. Insert the USB storage medium on the right side of the ControlCOCKPIT.
ð You can now use one of the programmes saved on the USB storage medium.
2. Press the activation key on the left of the Prog display.
ð The display is enlarged and the Select entry is automatically highlighted.
ð The programmes available for activation are shown on the right.
ð The programme currently available for use – in this example Test 012 – is
highlighted in orange.
3. Press the confirmation key to access the Select function.
ð All available programmes are displayed, including the ones saved on the USB data
storage medium (identified by the USB symbol ).
ð The programme currently available for use is highlighted in orange.
4. With the turn control, select the programme you want to use.
5. Confirm the selection by pressing the confirmation key.
ð The programme is now loaded, as indicated by the progress display.