Menu Mode
49 D53182 02/2024
Example: Temperature deviation should be corrected
1. Press the activation key to the right of the CALIB display.
ð The display is enlarged and the temperature adjustment option is automatically
highlighted.
2. Press the confirmation key repeatedly, until the calibration temperature Cal2 is
highlighted.
3. With the turn control, set the calibration temperature Cal2 to the specified
temperature.
4. Save the setting by pressing the confirmation key.
ð The corresponding calibration correction value is automatically highlighted.
5. Set the calibration correction value to 0.0 K.
6. Save the setting by pressing the confirmation key.
7. Position the sensor of a calibrated reference instrument centrally in the working
chamber of the appliance.
8. Close the door.
9. In manual mode, adjust the setpoint temperature.