Menu Mode
46 D53182 02/2024
i
If the temperature goes outside the limits of the tolerance band in setpoint-dependent
mode, the timer will be interrupted and only resumed once the temperature is within this
band once again.
See also
2 Digital Backwards Counter [}31]
8.3.6 Shelf Type (Grid or metal plate)
Here, you have to set the type of shelf (grid or metal panel) used. The selection Metal
panel enables you to adjust the closed-loop control function to the different air flow
characteristics in the chamber when using optional shelves instead of the grids provided
as standard.
8.3.7 Balance
For appliances of the size 55 and above, application-specific correction of the heat output
distribution (balance) between the upper and lower heating groups is possible. The setting
range is from –50% to +50%.
Distribution of the heating/cooling power (example):
The –20% (left) setting causes the lower heating elements to work at 20% less power than
the upper ones. The +30% (right) setting causes the lower heating elements to work at 30%
more power than the upper ones. The 0% setting restores the default distribution.