Menu Mode
53 D53182 02/2024
8.9 USER ID
8.9.1 Description
With the USER ID function, you can lock the settings of individual (e.g. temperature) or all
parameters, so that they cannot be changed at the appliance by accident or by
unauthorised persons.
You can also lock adjustment options in Menu mode (e.g. adjustment of date and time
settings).
■ If adjustment options are locked, this is indicated by the lock symbol in the respective
display.
i
USER ID data is entered in the AtmoCONTROL software and saved on the USB storage
medium. The USB storage medium is thus acting as a key: Parameters can only be locked
or unlocked if it is inserted.
A description of how to create a USER ID in AtmoCONTROL is provided in the separate
AtmoCONTROL manual.
8.9.2 USER ID Activation and Deactivation
1. Insert the USB storage medium with the USER ID data into the USB port on the right of
the ControlCOCKPIT.
2. Press the activation key on the right side of the USER-ID display.
ð The display is enlarged and the entry Activate automatically highlighted.
3. Confirm the activation by pressing the confirmation key.
ð The new USER ID data are transferred from the USB storage medium and
activated.
ð As soon as activation is complete, a check mark appears in front of the
corresponding entry.
4. Remove the USB storage medium.
ð Locked parameters are indicated by the lock symbol on the respective display.
i
To unlock the appliance,
■ insert the USB storage medium,
■ activate the USER ID display
■ and select the Deactivate entry.