80
AQUALAB 3
Discard used KIMWIPES strips frequently during the cleaning process to get each
component clean.
NOTE: Use the KIMWIPES strips supplied in the cleaning kit rather than facial tissue, paper towels, or
commercial swabs. KIMWIPES strips are specially made for lab work. They do not contain adhesives or
other contaminants and are made from low-lint fibers that will not scratch delicate surfaces when wet.
Figure98 View of inside of removed sensor block
8. With additional moistened KIMWIPES strips, clean the remaining surfaces of the sensor
block without wiping the KIMWIPES strips back over the sensors.
9. Moisten a fresh KIMWIPES strip wrapped around the cleaning swab with deionized water.
10. Clean the sensors and sensor block, following the same procedure outlined in step7. To
avoid cross-contamination, do not dip used KIMWIPES strips into clean distilled water.
11. Use fresh dry KIMWIPES strips wrapped around the cleaning swab to remove moisture
on the sensors and in the sensor block.