EasyManua.ls Logo

Metrohm 780 - Page 35

Metrohm 780
177 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Short operating tutorial
780/781 pH/Ion Meter, Manual
25
2
Start calibration with first standard
Immerse electrodes in the standard with the lowest F
-
concen-
tration
Start calibration with
<CAL>
If no temperature sensor is connected:
enter the temperature and exit with
<ENTER>
Confirm standard concentration with
<ENTER>
or if neces-
sarycorrect it
Wait for measurement of the potential
3
Continue calibration with increasing standard con-
centrations
When the request 'Change std.
<ENTER>
' appears, remove
the electrodes from the first standard, rinse them with water
and blot them gently with a clean lint-free tissue
Immerse electrodes in the next standard and continue the cal-
ibration sequence with
<ENTER>
In each case conform the standard concentration with
<ENTER>
or if necessary correct it
Repeat procedure for each remaining standard
4
Result
Wait for potential measurement, result will be displayed:
After 30 s the pH/Ion Meter will switch back automatically to
the measurement display. This can also be done immediately
with
<ENTER>
.
CAL
ENTER
ENTER

Table of Contents

Related product manuals