EasyManua.ls Logo

Metrohm 883 Basic IC plus - Transport Locking Screws

Metrohm 883 Basic IC plus
107 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.6 Installations on the rear of the instrument
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
883 Basic IC plus
3
Cable feed-through
For feeding through the detector cable.
4
Detector socket
For connecting the conductivity detector
(see chapter 3.15, page 43). Labeled
Detector.
5
Capillary feed-throughs
For feeding through the capillaries out of the
detector chamber.
CAUTION
The instrument must be switched off when connecting a detector.
1
Checking whether the instrument is switched off
If not, switch off the instrument.
2
Removing the back panel
Unscrew the knurled screws (4-1) on the back panel.
Remove the back panel (4-2).
3
Positioning the detector
Position the detector on the support surface intended for this pur-
pose and slide it right up to the front.
4
Replacing the back panel
Insert the detector cable in the cable feed-through (4-3) on the
back panel (4-2).
Replace the back panel (4-2) and tighten the knurled screws
(4-1).
5
Connecting the detector
Connect the detector cable to the detector socket (4-4).
3.6.2 Transport locking screws
To avoid damage to the high-pressure pump drive during transport, the
pump is secured with transport locking screws. These are located at the
rear of the instrument and labeled with Transport security screws
(2-5).
Remove these transport locking screws before the initial start-up.
Accessories
For this step you need:

Table of Contents

Related product manuals