4 Operation and maintenance
883 Basic IC plus
■■■■■■■■
73
Connecting the entry and exit for the high-pressure pump
1
■ Remove the stopper from the coupling. Tighten the coupling to
the pressure screw located on the pump head inlet capillary (1).
■ Reconnect the pump head outlet capillary to the pump head out-
let (2).
4.7 Servicing the inline filter
NOTICE
You can find a video sequence for this task in the Multimedia Guide IC
Maintenance or on the Internet at http://ic-help.metrohm.com/.
Maintenance interval
The filter must be replaced at least every 3 months; it may need to be
replaced more frequently, depending on the application.
Accessories For this task, you need the following accessories:
■ Two adjustable wrenches (6.2621.000) from the accessory kit: Vario/
Flex Basic (6.5000.000)
■ Tweezers
■ A new filter from the packaging (6.2821.130)