3 Installation
883 Basic IC plus
■■■■■■■■
47
3.18 Initial start-up
Even before the guard column and separation column are installed, the
entire system must be completely rinsed with eluent for the first time.
Rinsing the IC system
CAUTION
The separation column and the guard column are not permitted to be
installed at the time of the initial start-up.
Make sure that a coupling (6.2744.040) is being used instead of the
columns.
1
Preparing the software
Start the MagIC Net computer program.
Open the Equilibration tab in MagIC Net: Work-
place Run Equilibration.
Import (or create) a suitable method.
Also see: MagIC Net Tutorial and online help.
2
Preparing the instrument
Ensure that the eluent aspiration tubing is immersed in the eluent
and that there is enough eluent in the eluent bottle.
STREAM method (recommended): Ensure that the aspira-
tion tubing for the regeneration solution is immersed in the
solution and that there is enough solution. Check whether
the detector outlet capillary is connected to the Metrohm
Suppressor Module (MSM)'s inlet capillary for rinsing solu-
tion (labeled rinsing solution).
Alternatively, if two bottles are being used for rinsing and
regeneration: Ensure that the aspiration tubings for the aux-
iliary solutions (regeneration solution and rinsing solution)
are immersed in the respective solutions and that there is
enough solution in both bottles. Check whether the detec-
tor outlet capillary is guided into the waste container or is
connected with the waste collector.
Switch on the instrument.
MagIC Net detects the instrument and all of its modules.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 883 Basic IC plus and is the answer not in the manual?

Metrohm 883 Basic IC plus Specifications

General IconGeneral
BrandMetrohm
Model883 Basic IC plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals