3.12 Injection valve
■■■■■■■■■■■■■■■■■■■■■■
30
■■■■■■■■
883 Basic IC plus
Figure 12 Exchanging the sample loop
1
Pressure screw
Fastened to Port 6.
2
Sample loop
3
Pressure screw
Fastened to Port 3.
Exchanging the sample loop
NOTICE
Observe a dead-volume-free installation of the sample loop (see "Creat-
ing dead-volume-free capillary connections", page 16).
1
Removing the existing sample loop
Loosen the pressure screws (6.2744.010) at Port 3 and Port 6.
Remove the sample loop.
2
Installing a new sample loop
Fasten one end of the sample loop to Port 3 using a PEEK pressure
screw (6.2744.010).
Use the second PEEK pressure screw (6.2744.010) to fasten the
other end of the sample loop to Port 6.