3.14 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
883 Basic IC plus
2
Connect the capillary labeled rinsing solution to the outlet of the
pump tubing using a pressure screw (6.2744.070).
3
Connect the PTFE capillary from the rinsing solution bottle to the inlet
of the pump tubing.
3.14 Peristaltic pump
3.14.1 Installing the peristaltic pump
Installing the pump tubing
Pump tubing can differ in terms of material, diameter and thus flow rate.
Different pump tubing is used depending on the application.
Table 2
Pump tubings
Order num-
ber
Name Material Inner diame-
ter
Use
6.1826.310 Pump tubing LFL (orange/
green), 3 stoppers
PVC (Tygon
®
) 0.38 mm Pump tubing for bromate deter-
mination using the triiodide
method.
6.1826.320 Pump tubing LFL (orange/
yellow), 3 stoppers
PVC (Tygon
®
) 0.48 mm For acceptor solutions for Inline
Dialysis and for Inline Ultrafiltra-
tion.
6.1826.330 Pump tubing LFL (orange/
white), 3 stoppers
PVC (Tygon
®
) 0.64 mm No special applications.
6.1826.340 Pump tubing LFL (black/
black), 3 stoppers
PVC (Tygon
®
) 0.76 mm For sample solution in Inline Dial-
ysis.
6.1826.360 Pump tubing LFL (white/
white), 3 stoppers
PVC (Tygon
®
) 1.02 mm For sample transfer.
6.1826.380 Pump tubing LFL (gray/
gray), 3 stoppers
PVC (Tygon
®
) 1.25 mm For Inline Dilution.
6.1826.390 Pump tubing LFL (yellow/
yellow), 3 stoppers
PVC (Tygon
®
) 1.37 mm For sample solution in Inline
Ultrafiltration.
6.1826.420 Pump tubing PharMed
®
(orange/yellow), 3 stoppers
Ismaprene 0.51 mm For suppressor solutions.
Selecting the pump tubing and adapter
1
Select pump tubing suitable for the application (see table 2, page
38).