3 Installation
883 Basic IC plus
■■■■■■■■
37
3.13.3.3 Connecting the rinsing solution
There are 2 options for rinsing the suppressor:
Rinsing solution via STREAM (recommended)
Use the eluent from the conductivity detector as rinsing solution.
Rinsing solution via peristaltic pump
Prepare the rinsing solution in a separate bottle and transport with the
peristaltic pump.
The rinsing solution is connected to the rinsing solution capillary.
Connecting the rinsing solution inlet with STREAM
1
Connect the detector outlet capillary and the capillary labeled rinsing
solution to each other using a coupling (6.2744.040) and two pres-
sure screws (6.2744.070).
Connecting the rinsing solution inlet to the peristaltic pump
For this step, you need the following accessories:
Accessory kit: Flex/Vario: ChS (6.5000.030)
Pump tubing (6.1826.420)
Tubing olive with filter and locking nut (6.2744.180)
Tubing olive (6.2744.034)
Tubing cartridge of the peristaltic pump
1
Prepare a tubing cartridge of the peristaltic pump for the regenera-
tion solution (see chapter 3.14.1, page 38).