2 Installation
Titrando Installation Instructions 17
"Status" LED Dosing device operating status
off There is no exchange unit at-
tached.
constant illumination The Titrando is ready for dosing
or titrating. The exchange unit
has been attached correctly and
recognized and is now in the
change position, i. e. the ex-
change unit can be removed.
blinks slowly The Titrando is currently dosing
or filling or the exchange unit is
not in the change position.
An intelligent 806 Exchange unit
has been attached and the data
on the built-in data chip is cur-
rently being read out or written.
blinks rapidly Dosing drive error (see Sec-
tion 3, Troubleshooting)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm Titrando and is the answer not in the manual?

Metrohm Titrando Specifications

General IconGeneral
mV Resolution0.1 mV
Temperature Resolution0.1 °C
ApplicationsAcid-base, redox, complexometric, precipitation, Karl Fischer titration
Measuring inputspH, mV, temperature
Stirrer controlIntegrated stirrer control
Softwaretiamo
Titration MethodsMET (Monotonic Equivalence Titration), SET (Fixed Equivalence Titration)
Dosing ResolutionVaries depending on the Dosing device used
Minimum Dosing VolumeVaries depending on the Dosing device used
pH Resolution0.001
InterfaceUSB
Power Supply100 – 240 V AC (depending on the model)
DimensionsVaries depending on the model
WeightVaries depending on the model