AccuPyc II 1340 Appendix A
Oct 2013 A-1
A. ERROR MESSAGES
This appendix contains an alphabetical listing of the messages that may appear in the display
of the pycnometer.
Operational and report error messages generated during an analysis or calibration are placed in
a queue, along with report data. Error messages are indicated with an asterisk at the Reload
prompt after the operation is completed.
Messages and data are placed in the following order:
Operational error messages
Report error messages
Report data
Press
CHOICE to cycle through the messages and to the data. If you wish to delete the message
after viewing its contents, press
CLEAR. If you aren’t clear as to why the message was
displayed, look it up in this appendix for a cause and a resolution before deleting. Messages
that are not cleared will remain in the queue until the next analysis or calibration is performed;
at that time, any remaining messages will be cleared automatically.
When you see NN/ZZ in a message listed below, NN = number of cycles completed and
ZZ = number of cycles requested.
Add Unit
Cause A: A new analysis module has been attached to the instrument where
previously a different module had been attached.
Action A: Press
Choice and select Yes to add the new analysis module.
Cause B: An incorrect serial number was read by the analysis module.
Action B: Press
Choice and select No to use the serial number and calibration
information stored in the command module. If this message displays
every time you turn on the instrument, contact your service
representative.
Unit[n] SN1234 10 cm3
X - O - X
Reload *
P = (current pressure)
An asterisk (*) indicates error message