AccuPyc II 1340 System Commands
Aug 2013 3-25
Calibrate
This command enables you to perform a full volume slope and offset calibration; enter a temperature
or pressure value obtained from an external reference device; copy to or load from a USB stick; and
reset the pressure calibration.
Press
ALT + . (decimal) to access the Calibrate function.
Enables you to choose the insert you are using.
Choices: None, 0.1 cm
3
(for 1 cm
3
unit)
None, 3.5 cm
3
, 1.0 cm
3
(for 10 cm
3
unit)
None, 35 cm
3
, 10.0 cm
3
(for 100 cm
3
unit)
None, 650 cm
3
, 1300 cm
3
(for 2000 cm
3
pycnometer)
Refer to MULTIVOLUME INSERTS, page J-1 for
additional information on MultiVolume inserts.
Press
CHOICE until the appropriate insert is displayed or
None if an insert is not being used, then press
ENTER.
Choose one of the following:
• Press
ENTER to start the analysis.
The analysis begins and operational status messages are
continually displayed during analysis.
• Press
Alt + CLEAR to cancel the analysis.
• Press
SAVE to store the data you entered and return to dis-
play mode.
Choices: Volume, Temperature, Pressure,
Copy to USB Stick, Load from USB Stick,
Reset Pressure Calibration
Press
CHOICE until the desired option is displayed:
Press
ENTER.
Unit[n] SN1234 10 cm3
Analyze
Chamber insert?
None
Unit[n] SN1234 10 cm3
Analyze
[ENTER] to start
[Escape] to cancel
Unit[n] SN1234 10 cm3
X - X - X
Calibration Type?
Volume