AccuPyc II 1340 Appendix E
Dec 2011 E-3
Connecting One or More Analysis Modules to a Controller
Some models of the AccuPyc require a Control module and an Analysis module as separate
units. For example, the Glove box system and the 350-cm
3
system. A Control module contains
a keypad and a display, but does not contain a sample compartment.
Attach an Analysis module to a Control module as follows:
:
1. Place the On/Off switch on the rear panel of the analyzer in the Off ( O ) position.
2. Locate the power adapter (in the analysis module kit) and connect external power supply
cable to each power adapter to be used to connect the analysis modules below the din
connector on the power adapter.
3. Connect a power cable to each external power supply.
4. Attach one end of the din cable (provided in the control module accessories kit) to the
connector labeled Module Connection on the rear panel of the single-unit system.
5. Attach the other end of the din cable to the separate power adapter.
Power
Adapter
Connect to Wall Outlet
POWER
INPUT
NEXT
Next
Unit
OUT
Control Module
Analysis
Module
External
Power Supply
Power Supply
External
Power Cable
IN
Din Cable
If you have the Windows version of the AccuPyc, exit the application before turning off
the analyzer.