Appendix E AccuPyc II 1340
E-2 Dec 2011
2. Connect one external power supply cable to the back of the single unit system and another
external power supply cable to each power adapter to be used to connect the analysis
modules.
3. Connect a power cable to each external power supply.
4. Attach one end of the din cable (provided in the analysis module accessories kit) to the
connector labeled Module Connection on the rear panel of the single-unit system.
5. Attach the other end of the din cable to the separate power adapter.
6. Attach the hardwired power adapter cable to the lower connector labeled Module
Connection on the rear panel of the analysis module. Be sure to use the lower connector
on the analysis module; the upper connector is for attaching another analysis module.
7. Plug the external power supply cables into electrical wall outlets.
8. Place the On/Off switch on the rear panel of the analyzer in the On ( | ) position.
9. Press
Alt + CHOICE + 2 to access the second Analysis module; Unit2 should display in
the upper left corner of the display:
10. Calibrate the Analysis module using the Calibration Kit shipped with the module. Refer to
Calibrating the Pycnometer in Chapter 2 for step-by-step instructions.
Unit2 SN1234 10 cm3
X - X - X
Reload
P = (pressure)
If you have the Windows version of the AccuPyc, Unit menus are added to the menu
bar for each analysis module you attach.