EasyManua.ls Logo

Micromeritics AccuPyc II 1340 - Specifications

Micromeritics AccuPyc II 1340
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications AccuPyc II 1340
1-8 Aug 2013
Specifications
The AccuPyc II 1340 Pycnometer has been tested to meet the specifications in the following table.
Characteristic Specification
Environment
Temperature: Stable between 15 and 35 ºC (59 to 96 ºF)
Temperature-controlled AccuPyc: temperature stability is dependent
upon specifications of the installed circulator. Recommended range: 15
to 50 ºC (59 to 122 ºF)
Humidity:
20 to 80% relative, non-condensing
Physical
Height:
17.9 cm (7.0 in.); analysis modules for 1-, 10-, and 100-cm
3
units, and
control module
25.9 cm (10.2 in.), 350 cm
3
analysis module
43 cm (17 in.), 2000 cm
3
analysis module
Width: 27.3 cm (10.7 in.), control module
22.2 cm (8.7 in.), analysis module
27 cm (10.6 in.), 2000 cm
3
analysis module
Depth: 36.2 cm (14.3 in.)
Weight:
9.3 kg (20.5 lbs), control/analysis unit (1-, 10-, and 100-cm
3
units)
7.9 kg (17.4 lbs), analysis module (1-, 10-, and 100-cm
3
units)
10.5 kg (23.2 lbs), analysis module (350-cm
3
unit)
3.6 kg (8.0 lbs), control module
26 kg (57.0 lbs), 2000 cm
3
analysis module
Electrical
Voltage: 90 to 264 VAC
Power: 30 VA
Frequency: 50 to 60 Hz

Table of Contents