AccuPyc II 1340 Appendix E
Dec 2011 E-1
E. MULTIPLE ANALYSIS MODULES
This Appendix provides information on attaching multiple analysis modules to a control mod-
ule. You can have up to six analysis modules for one control module. In most cases, a control
module also contains an analysis module in a single unit. Some systems are available only as a
control module and analysis module (two units); for example, the 350-cm
3
system and the
Glove Box system.
Connecting Modules to a Single Unit System
Additional analysis modules enable you to increase productivity and/or provide means for
analyzing different size samples concurrently. For example, you may wish to attach an addi-
tional 10-cm
3
analysis module to an existing 10-cm
3
system, or a 100-cm
3
analysis module to
an existing 10-cm
3
system.
To attach an additional analysis module to an existing single-unit system (controller and anal-
ysis modules combined):
Attach an Analysis module to a Control module as follows:
1. Place the On/Off switch on the rear panel of the analyzer in the Off ( O ) position.
Power
Adapter
Connect to Wall Outlet
POWER
INPUT
NEXT
IN
Next
Unit
OUT
Din
Cable
Analysis
Module
External
Power Supply
Power Supply
External
Power Cable
Single Unit
If you have the Windows version of the AccuPyc, exit the application before turning off
the analyzer.