System Commands AccuPyc II 1340
3-26 Aug 2013
Volume
Before proceeding, place an empty cup into the sample chamber for Cal1 (volume offset). If using an
insert, place the insert and its appropriate sample cup into the sample chamber. (Refer to Installing
and Removing Inserts and Sample Cups, page J-2.)
Press
ENTER to accept Volume (default) and display the
next prompt.
Press
CHOICE until the appropriate insert is displayed, then
press
ENTER.
Choices: None, 0.1 cm
3
(for 1 cm
3
pycnometer)
None, 3.5 cm
3
, 1.0 cm
3
(for 10 cm
3
pycnometer)
None, 35 cm
3
, 10 cm
3
(for 100 cm
3
pycnometer)
None, 650 cm
3
, 1300 cm
3
(for 2000 cm
3
pycnometer)
Yes must be selected for the Request Insert prompt in
Report Options for this prompt to display.
Enter the volume of the calibration standard(s) you plan to
use in the calibration (located on the outside of the
calibration standards case).
Range: 0.1 to 999.0000 cm
3
Some pycnometers require and are shipped with multiple
standards; in this instance, the value recorded on the case is
the sum for all standards and is the one you should enter at
this prompt. Press
ENTER.
Press
ENTER to begin the calibration or Alt + CLEAR to
cancel the calibration.
The volume offset is calibrated first (Cal1). When the first
calibration is complete, the pycnometer beeps three times
and the next prompt is displayed.
Unit[n] SN1234 10 cm3
X - X - X
Calibration Type?
Volume
Unit[n] SN1234 10 cm3
Calibrate Volume
Chamber insert?
None
Unit[n] SN1234 10 cm3
Calibrate Volume
Volume of Cal Std:
1.0000 cm3
Unit[n] SN1234 10 cm3
Calibrate Volume
[ENTER] to start
[Escape] to cancel