AccuPyc II 1340 Appendix A
Oct 2013 A-3
ANLSERR: Underrange
NN/ZZ cycles completed
Automatic operation
has been canceled
CAL_ERR: Reset Pressure Cal. ?
CAL_ERR: Temperature
Calibration Failed
Cause: A pressure underrange occurred during an analysis because an error
occurred in the pressure measurement electronics.
Action: Call a Micromeritics Service Representative.
Cause: The automatic operation has been canceled by the user.
Action: Wait for the termination process to complete or end the termination
process by pressing
Alt + CLEAR two more times.
Cause: The calibration information could not be read from the command or
analysis module.
Action:
Choose
Yes and press Enter to reset the pressure and temperature
calibration to nominal values. Resetting to nominal may restore
normal readings but a proper calibration should be performed by a
service technician. See the section Reset Pressure Calibration in the
System Commands section.
Action:
Choose
No to avoid resetting the calibration. The instrument will be
unresponsive.
Cause: The temperature calibration was unsuccessful.
Action: Contact your service representative.