AccuPyc II 1340 Appendix A
Oct 2013 A-11
USR_ERR:
No data to review
USR_ERR: Number of
cycles must be = 5
USR_ERR: Out of range
USR_ERR: Pressure overrange
Valve failure
[ENTER] to retry.
Cause: You tried to review data for an automatic operation when there were no
data to review.
Action: Abandon request.
Cause: You tried to enable run precision without increasing the number of runs
to at least five.
Action: Increase the number of runs to five or abandon request.
Cause: You tried to enter a value that is out of the valid range.
Action: Enter a value in the specified range (refer to Chapter 4).
Cause: A pressure overrange occurred but was left uncorrected.
Action: Return to manual mode and vent the system.
Cause: One (or more) of the system valves failed to operate when the
instrument was initialized.
Action: Press
ENTER to retry. If still unsucessful, contact your service
representative.