6. Press CHOICE until the average measured volume is displayed; record the value.
7. Press
Alt + 2 to access Setup.
8. Press
CHOICE until Calibration Data is displayed.
9. Press
ENTER; the Chamber Insert prompt is displayed.
10. Ensure that None is selected, then press
ENTER to display the Cell Volume prompt.
11. Subtract the average volume of the 0.1-cm
3
insert/cup (recorded in Step 6) from the cell
volume displayed in the prompt; record this value.
12. Press
ENTER to display the Expansion Volume prompt; record the value displayed.
13. Record the value displayed for the expansion volume, then press
ENTER. This returns you
to the Calibration Data prompt.
14. Press
ENTER; the Chamber Insert prompt is displayed.
15. Press
CHOICE until 0.1 cm
3
is displayed, then press Enter to display the Cell Volume
prompt. Enter the value you recorded in Step 11.
16. Press
ENTER to display the Expansion Volume prompt; enter the value you recorded in
Step 12.
17. Press
SAVE to save the information you entered and return to the Reload prompt.
The pycnometer is now ready for analyses using the 0.1-cm
3
insert/cup.
Unit[n] SN1234 10 cm3
Set-up Type?
Calibration Data