Appendix K AccuPyc II 1340
K-2 Apr 07
5. Remove the protective caps from the ends of the gas entrance tubing, then remove the
ferrules from the upper end of the tubing.
6. Remove the nut from the center port and slide it onto the end of the gas entrance tubing,
then replace the front and rear ferrules. Be sure the ferrules are oriented as shown below.
7. Insert the tubing into the center port of the Multigas assembly; use the nut to secure the
tubing to the assembly.
8. Insert the other end of the tubing into the gas inlet on the rear panel of the analyzer; use the
self-contained swage nut to secure the tubing to the inlet.
Ferrules
Swage nut
Ferrules
Swage nut