System Commands AccuPyc II 1340
3-12 Aug 2013
pycnometer is shipped with default values; however, you may wish to modify them to meet your labo-
ratory requirements.
Press
ENTER to accept Analysis Parameters and display the
next prompt.
Enter the number of purges to be performed and press
ENTER.
Range: 0 to 999
Purging cleans the sample cell and expansion chambers
before an analysis begins. The greater the number of purges,
the cleaner the sample will be when analyzed.
For calibration, 10 is sufficient.
Enter the fill pressure and press
ENTER. For most
applications, the default of 19.500 psig is adequate.
Range: 0 to 19.850 psig
0 to 136.86 kPag
Typically, the greater the fill pressure, the easier it is to
measure the volume precisely. However, a lower pressure
may be desirable for some samples.
Enter the number of cycles to be performed and press
ENTER. A cycle is a series of functions which equal to a
single volume measurement.
Range: 1 to 999
For calibration, 10 is sufficient.
Enter the fill pressure and press
ENTER. For most
applications, the default of 19.500 psig is adequate.
Range: 0 to 19.850 psig
0 to 136.86 kPag
Unit[n] SN1234 10 cm3
Set-up Type?
Analysis Parameters
Unit[n] SN1234 10 cm3
Analysis Parameters
Number of Purges:
10
Unit[n] SN1234 10 cm3
Analysis Parameters
Purge Fill Pressure:
19.500 psig
Unit[n] SN1234 10 cm3
Analysis Parameters
Number of cycles:
10
Unit[n] SN1234 10 cm3
Analysis Parameters
Cycle Fill Pressure:
19.500 psig