AccuPyc II 1340 System Commands
Aug 2013 3-29
Reset Pressure Calibration
This option can be used when you suspect the pressure is not being reported correctly by the analyzer.
For example: a zero, negative, or unreasonable high reading is being consistently reported. Resetting to
nominal may restore normal pressure readings but a proper calibration should be performed.
Enter the serial number, then press
ENTER.
These prompts appear in succession.
Remove the USB stick; the system automatically returns to
the Reload prompt.
Press
CHOICE until the desired option is displayed:
Press
ENTER.
This resets the pressure calibration to nominal values;
temperature and volume calibrations are not affected.
Nominal values may be sufficient for your application.
However, you should schedule a proper pressure calibration
as soon as possible. Press
CHOICE.
Choices: Yes, No
Select Yes to reset the calibration to nominal values. Contact
your service representative and schedule a proper pressure
calibration since data accuracy may be compromised.
Select No if you wish to contact your service representative
to schedule a proper pressure calibration.
Press
ENTER.
Unit[n] SN1234 10 cm3
X - X - X
Load cal. for SN:
Unit[n] SN1234 10 cm3
X - X - X
Calibration for
SN (nnnn) loaded
Unit[n] SN1234 10 cm3
X - X - X
Remove USB stick
Unit[n] SN1234 10 cm3
X - X - X
Calibration Type?
Reset Pressure Cal.
Unit[n] SN1234 10 cm3
X - X - X
Reset Pressure Cal.?
No