Maintenance AccuPyc II 1340
5-6 Aug 2013
5. Grasp the O-ring with the two greased fingers.
6. With the other hand, pull the O-ring through the greased fingers until the entire O-ring is
greased.
7. Place the O-ring back into the groove on the cap and, with the greased index finger,
gently press it back into position. After the O-ring is fully seated, grease the groove with
your index finger without applying additional grease.
8. Replace the chamber cap.