Pre Installation
Installation Requirements
Pretreated water
Requirements
Type of pretreated water Elix system, RO, distilled or DI water
Temperature
5 to 35 °C
Environmental
Requirements
Indoor Use Only
Ambient Storage
Temperature
5 °C < T < 40 °C
Ambient Operating
Temperature
5 °C < T < 40 °C
Relative Humidity
Should not exceed 80% for temperatures below 31 °C
Should stay within 50% to 80% between 31 °C and
40 °C.
Altitude < 3000 metres
Installation Category II
Pollution Degree 2
Optional Equipment You May Need
Wall Mounting
Bracket
The Catalogue Number for the Wall Mounting Bracket is WMBSMT002. The mounting
hardware for attaching the bracket to a wall is not included and must be supplied.
Unpacking the Simplicity System− What’s Inside?
Open the Simplicity System Shipping Box. Use the checklist included in the Accessories
Bag to make sure all items were shipped and are accounted for. It is highly suggested
to become familiar with the items that are shipped since these will be used in the
Installation section of this manual.
Contact Millipore SAS if an item is missing.