25
Maintenance
Maintenance Schedule
What to do When? How to?
SimpliPak
Cartridge
Replacement
When the Pack
Alarm display is
blinking.
See Section How
to Replace the
SimpliPak
cartridge
When the system
resistivity display
is blinking.
Final Filter
Replacement
The SimFilter Final Filter is replaced
when the SimpliPak cartridge is
replaced or when the Product Water
flowrate drops.
The BioPak cartridge should be replaced
every 90 days.
See Section How
to Replace the
Final Filter
Tank Level
Calibration
Periodically.
See Section How
to Calibrate the
Tank Level (C03)
UV Lamp
Replacemen
t and UV
Timer Reset
When the UV
Lamp Alarm
display is
blinking.
See Section How
to Replace the UV
Lamp (UV System
only) and How to
View or Reset the
UV Lamp Timer
(C04)