Maintenance
36
How to Reset the UV Lamp Timer
1
Press the Main and “+“ Buttons together
to enter the menu. The Display will show
C01.
2
Press the Main Button 3 times. The Display
will show C04 and “0” days left on the UV
Timer.
3
Press the “+” and “-“ Buttons together.
This will reset the UV Timer to 500 (days).
The Display will exit the menu.
The UV Timer has been reset.