P907 PROGRAMMER INTERFACE
to activate the programmer interface (two way infrared
communications) test.
On successful test completion, "PASS" is displayed. Otherwise, "FAIL"
is displayed.
P910 RELAY
Access this parameter to display the current state of the specified relay.
to alter the relay state (energized / de-energized) or enter
the value corresponding to the specific relay state desired.
values: 0 = de-energized
1 = energized
P911 mA OUTPUT VALUE
Access this parameter to display the current value of the mA output.
Additionally, this feature may be used to enter a desired value. The mA
output immediately assumes the value entered regardless of any
restrictions programmed.
values: 0.000 to 25.00
P912 TRANSDUCER TEMPERATURE
Use this feature to display the temperature in ° C (as monitored by the
connected transducer). "Err" is displayed if the transducer is not equipped
with an internal temperature sensor.
values: -50 to 150
P913 SENSOR TEMPERATURE
Access this parameter to display the temperature in ° C (as monitored by
the TS-3). "OPEn" is displayed if a TS-3 is not connected.
values: -50 to 150
P914 mA INPUT
Use this feature to display the mA input value (in mA).
values: 0.000 to 22.00
PARAMETERS
PL-507 145