OCM PARAMETERS (P600 - P607, P610, P611, P620, P621)
If The HYDRO+ is used to monitor open channel flow, alter the following parameters
as required.
The HYDRO+ calculates "head" as referenced to Empty (P006) or OCM Zero Offset
(P605), when Operation is set for "OCM" (P001 = 6). Flowrate, based on head
(at the "point of measure" specified by the PMD fabricator) is also calculated.
P600 (G) PRIMARY MEASURING DEVICE
Enter the type of primary measuring device (PMD) used.
If Operation is not set for "OCM" (P001 = 6), this value is preset to "0". If
Operation is set for "OCM", this value is preset to "1".
The HYDRO+ is preprogrammed with the flow constant for common PMD
flow calculations. If the PMD used is not specifically identified, select the
appropriate Universal Flow Calculation.
Associated parameters (as indicated below), Max Head (P603), Max Flow
(P604), and Min Head (P605) may be scroll accessed.
values: 0 = off (no calculation)
1 = exponential (see P601)
2 = Palmer-Bowlus Flume (see P602)
3 = H-Flume (see P602)
4 = Universal Linear Flow Calculation (see P610, P611)
5 = Universal Curved Flow Calculation (see P610, P611)
6 = BS - 3680 / ISO 4359 Rectangular Flume
7 = BS - 3680 / ISO 1438/1 Thin Plate V-notch Weir
PARAMETERS
PL-507 105

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Milltronics HydroRanger Plus and is the answer not in the manual?

Milltronics HydroRanger Plus Specifications

General IconGeneral
BrandMilltronics
ModelHydroRanger Plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals