TEST PARAMETERS (P900 to P913)
Test Parameters are intended for use by Milltronics Service personnel.
P900 (V) SOFTWARE REV. #
View the EPROM Rev. # without removing the enclosure lid.
values: 00.00 to 99.99
P901 (V) MEMORY
to activate the HYDRO+ memory test.
values: PASS (memory test successful)
F1 = RAM F3 = EEPROM
F2 = NOVRAM F4 = EPROM
P902 (V) WATCHDOG
to reset the microprocessor.
On successful completion (10 seconds) the RUN mode is entered.
P903 (V) DISPLAY
to activate the display test.
All LCD segments and symbols are temporarily displayed.
P904 (G) KEYPAD
Press each keypad key in the following sequence:
As each key is pressed, the associated keypad number is displayed. On
successful test completion, "PASS" is displayed. "FAIL" is displayed if a key
is pressed out of sequence or the programmer keypad malfunctions.
P905 TRANSMIT PULSE
to supply repeated transmit pulses to the transducer and / or
view the transducer operating frequency (automatically altered by (P004)
Transducer) for the Point Number displayed.
This feature may be used to monitor the transmit pulse with an oscilloscope
connected to the transducer terminals.
values: 10.00 to 60.00
PL-507 144

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Milltronics HydroRanger Plus and is the answer not in the manual?

Milltronics HydroRanger Plus Specifications

General IconGeneral
BrandMilltronics
ModelHydroRanger Plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals