P062 OFFSET READING
Enter the value to be added to the Reading, (before display).
This feature is preset to 0.000, (no offset).
e.g. To reference the current level to sea level, enter the distance in Units
(P005), between Empty (P006) and sea level. (Enter a negative value
if Empty is below sea level.)
This feature affects the Reading only.
(Relays and mA outputs are not affected).
values: -999 to 9999
vessel
display offset
(P062)
Empty
(P006)
sea
level
PL-507 68