EasyManuals Logo

Mindray A3 Service Manual

Mindray A3
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #117 background imageLoading...
Page #117 background image
A5/A3â„¢ Service Manual 046-001141-00 2 - 31
Installation Guide Functional Tests
FIGURE 2-44 O2 Sensor Calibration Successful
2.3.3.2 100% O2 Calibration
NOTE: 100% O2 calibration must be performed in standby mode.
NOTE: 100% O2 calibration can be performed only after a successful 21% O2
calibration.
NOTE: IMake sure that the manual bag is in position in manual mode.
Otherwise, put the Manual/Auto lever to Auto position.
Follow these steps to calibrate O2 sensor at 100% O2:
1. Enter Standby.
2. Select Setup > System > Calibration > O2 Sensor or Setup > Service > Calibration > O2 Sensor.
The System and Service tabs require passwords and shows both 21% and 100% O2 Sensor
calibration. The calibration screen shown below is displayed when 100% is selected. Set up the
machine as per the instructions on the screen and select Next.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A3 and is the answer not in the manual?

Mindray A3 Specifications

General IconGeneral
BrandMindray
ModelA3
CategoryMedical Equipment
LanguageEnglish

Related product manuals