EasyManua.ls Logo

Mindray A3 - Page 117

Mindray A3
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A5/A3™ Service Manual 046-001141-00 2 - 31
Installation Guide Functional Tests
FIGURE 2-44 O2 Sensor Calibration Successful
2.3.3.2 100% O2 Calibration
NOTE: 100% O2 calibration must be performed in standby mode.
NOTE: 100% O2 calibration can be performed only after a successful 21% O2
calibration.
NOTE: IMake sure that the manual bag is in position in manual mode.
Otherwise, put the Manual/Auto lever to Auto position.
Follow these steps to calibrate O2 sensor at 100% O2:
1. Enter Standby.
2. Select Setup > System > Calibration > O2 Sensor or Setup > Service > Calibration > O2 Sensor.
The System and Service tabs require passwords and shows both 21% and 100% O2 Sensor
calibration. The calibration screen shown below is displayed when 100% is selected. Set up the
machine as per the instructions on the screen and select Next.

Table of Contents

Related product manuals